Liquiline System CA80AL
Operation
Hauser
75
Menu/Setup/Sample preparation
Function
Options
Info
Pump time
Selection
10 to 20 s
Factory setting
10 s
Ratio between the pumping duration and the
pumping interval of the peristaltic pump. This
affects the volume of the pumped sample.
Pump pause
Selection
20 to 50 s
Factory setting
30 s
Pressurized air cleaning
(CAT820, CAT860)
Compressed air
(CAT820)
Selection
• Available
• Not available
Factory setting
Depends on the device
version
Settings for sample preparation with or
without compressed air cleaning.
In the event of retrofitting, the compressed air
can be activated here.
Cleaning mode
Selection
• On
• Off
Factory setting
On
Activates or deactivates automatic
compressed air cleaning of the pump-to-filter
hose and of the filter.
Cleaning interval
Selection
0:30 to 4:00 HH:MM
Factory setting
2:00
Cleaning interval of the automatic compressed
air cleaning system
Cleaning duration
Selection
10 to 60 s
Factory setting
30 s
Cleaning duration of the automatic
compressed air cleaning system
Heatings
(CAT820, CAT860)
Cabinet
Selection
• Available
• Not available
Factory setting
Depends on the device
version
In the event of retrofitting, the heating can be
activated here.
Hose filter
Selection
• Available
• Not available
Factory setting
Depends on the device
version
In the event of retrofitting, the heating can be
activated here.
Hose analyzer
Selection
• Available
• Not available
Factory setting
Depends on the device
version
In the event of retrofitting, the heating can be
activated here.
Diagnostics settings
Limit change filter
(CAT820, CAT860)
Function
Selection
• On
• Off
Factory setting
Off