Liquiline System CA80AL
Maintenance
Hauser
145
Replacing bottles
1. Replace the removed bottles with fresh reagent , standard solution or cleaner .
Reagents must be prepared according to the instructions for mixing a reagent.
2. Reconnect the hoses to the Liquid Manager.
3.
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle
selection
must be selected.
4. Select all the bottles that you have replaced and press
OK
to confirm.
5.
Confirm bottles inserted
must be selected.
6. If bottle level monitoring is enabled (
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Diagnostics settings
/
Bottles
), it can be reset under
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Reset filling levels
.
7. Zero point calibration of the system must be performed after a replacement routine
8. After the zero point calibration, select
Menu
/
Operation
/
Manual operation
/
Determine calibration factor
.
9. Go back to
MODE
/
Continue automatic mode
or
MODE
/
Start automatic mode
to
start normal measurement.
12.3.1 Shelf life of the reagents
Ready-to-prepare reagent
6 months
Ready-to-use reagent
6 weeks
Under unfavorable ambient conditions, the shelf life of the reagents can expire even
before the reagents are consumed.
12.4 Carrying out zero point calibration
Replacing standard calibration solution with zero standard
1. Press the
MODE
soft key and select the
Manual mode
.
The display shows
Current mode
-
Manual
.
Wait until all operations have stopped. Ongoing operations can be stopped under
Menu
/
Operation
/
Manual operation
.
2. Remove the bottles containing standard S1 by removing the hose from the bottle. Use
a paper towel to catch any reagents that may escape. This prevents any
contamination of the zero standard.
3. Replace the standard with zero standard.
4. Connect the zero standard to the Liquid Manager, connection S1.
5. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
6. Select
Standard S1
and press
OK
to confirm.
7. Select
Confirm bottles inserted
.
Determining the zero point
1. Call up:
Menu
/
Operation
/
Manual operation
/
Determine zero point
.
2. Carry out zero point calibration.
3. Following successful calibration, the following message appears: "Do you want to
accept the calibration data for adjustment?". Press
OK
to confirm.