Operation
Liquiline System CA80AL
74
Hauser
Menu/Setup/Sample preparation
Function
Options
Info
Installation type
Selection
• Others
• Pressurized pipe
sampling
Factory setting
Others
Others:
e.g. Liquiline System CAT820 / CAT860
Pressurized pipe sampling:
Sample preparation that is installed in a
pressure pipe, e.g. Liquiline System CAT810.
If the installation type selected is
Others
:
Sample preparation 1
In the case of two-channel devices, sample
preparation 2 is also displayed.
Operating mode
Selection
• Controlled
• Independent
Factory setting
Controlled
Controlled:
Sample preparation which is controlled by the
CA80 analyzer,
e.g. Liquiline System CAT820 / CAT860
Independent:
No sample preparation that is controlled by
the Liquiline System CA80 analyzer, e.g.
Stamoclean CAT430. The supply of sample
must be guaranteed by the customer.
If the operating mode selected is
Controlled
:
Tag
Customized text,
32 characters
Cleaning start condition
(CAT860)
Selection
• Immediate
• Date/time
• Disabled
Immediate:
Cleaning starts immediately.
Date/time:
Cleaning starts at the selected date/time.
Disabled:
Cleaning is not active.
If the start condition selected is
Immediate
or
Date/time
:
Cleaning interval
(CAT860)
Selection
• 0-01 to 90-00 (DD-HH)
•
Factory setting
1-00
Compressed air or water cleaning for longer
filter maintenance intervals.
Cleaning with liquid cleaner.
If the start condition selected is
Date/time
:
Date
(CAT860)
Selection
• 01.01.1970 to
07.02.2106
DD.MM.YYYY
•
Factory setting
01.01.1970
Time
(CAT860)
Selection
• 00:00:00 to 23:59:59
HH:MM:SS
•
Factory setting
00:00:00
Residence time
(CAT860)
Selection
• 0:30 to 20:00 (MM:SS)
•
Factory setting
1:00
Reaction time of cleaner during the cleaning
Sampling pump interval
(CAT820, CAT860)