Liquiline System CA80AL
Operation
Hauser
73
Menu/Setup/Analyzer/Calibration
Function
Options
Info
Calibration factor
Read only
Relationship of the measured concentration to
the pre-specified concentration of the calibration
standard.
Settings
Nominal concentration
1.00 to 1000.00 (μg/l)
Factory setting
500.00 μg/l
For configuring the concentration of the
calibration standard solution.
Depends on the settings under
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Measuring value
/
Main value
Automatic cleaning
Selection
• Off
• On
Factory setting
Off
For specifying whether cleaning will take place
before every calibration/adjustment (only in the
automatic mode).
10.3.4 Cleaning
Additional cleaning can be adapted to the application conditions in this menu.
Menu/Setup/Analyzer/Cleaning
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
• Disabled
Factory setting
Disabled
Immediate
:Cleaning of the starts immediately.
Date/time
: Cleaning of the starts at the selected
date/time.
Disabled
: Additional cleaning is not enabled.
If the start condition selected is
Date/time
Date
01.01.1970 to 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24 h)
Cleaning interval
0-01 to 90-00 (DD-HH)
Factory setting
02-00
For configuring the time interval for (additional)
cleaning.
Next cleaning
Mode
=
Automatic
Read only
10.4 Sample preparation
The menu displayed depends on the connected sample preparation system. The menu
item does not appear if a Liquiline System CA80 analyzer without a collecting vessel is
connected. The Liquiline System CAT860 can only be operated with a Liquiline
System CA80 single-channel device.