Error messages
Solutions
Err 100:
Bad calibration
Refer to chapter "Calibration".
•
Err 101:
Absence or damage to the halogen
bulb
Place or change the bulb and check light intensity. If necessary calibrate the unit.
•
Err 103:
Check that the input voltage is indeed 115 V or 230 V,
Control the cooling liquid level.
•
•
Err 104:
Blocked water circuit
Check that the cord is not jammed or bent.
•
Err 105:
Cooling liquid temperature too high
Check the level of the cooling liquid,
Wait for the liquid to cool down before continuing the treatment,
Replace water in the reservoir with new distilled (or demineralized) water. Refer to chapter "Replacing the
cooling liquid".
•
•
•
Err 107:
Check the positioning of the reservoir.
•
Err 108:
Insufficient light power
Check that the light guide and the radiometer window are perfectly clean,
Calibrate the unit,
If lamp replacement indicator is still flashing, change the halogen bulb.
•
•
•
Err 126:
Check if one of the keys of the front panel is stuck. To release the key press all the keys one after the other.
•
Err 127:
Check that the cord is correctly screwed in,
Check that the handpiece button is not activated.
•
•
If an error continues:
Switch the unit off and then back on. If the error is still displayed, contact your after-sales service.
•
Errors other than those listed above
may be displayed on your screen
Switch the unit off and then back on. If the error is still displayed, contact your after-sales service.
•