31
Reference Manual
00809-0100-4853, Rev AB
Installation
January 2017
Installation
Figure 3-15. Rosemount 305 Manifold Styles
The Rosemount 304 comes in two basic styles: traditional (flange x flange and flange x pipe) and wafer.
The traditional manifold comes in 2-, 3-, and 5-valve configurations. The wafer manifold comes in three-
and five-valve configurations.
Figure 3-16. Rosemount 304 Manifold Styles
3.5.1
Rosemount 305 Integral Manifold installation procedure
To install a Rosemount 305 Integral Manifold to a Rosemount 3051SMV:
1. Inspect the PTFE SuperModule
™
O-rings. If the O-rings are undamaged, reusing them is
recommended. If the O-rings are damaged (e.g. nicks), replace them with new O-rings.
Note
If replacing the O-rings, be careful not to scratch or deface the O-ring grooves or the surface of the
isolating diaphragm when removing the damaged O-rings.
2. Install the Integral manifold on the SuperModule process connection. Use the four manifold bolts for
alignment. Finger tighten the bolts, then tighten the bolts incrementally in a cross pattern to final
torque value. See
for complete bolt installation information and torque values.
When fully tightened, the bolts should extend through the top of the SuperModule housing.
3. If the PTFE SuperModule O-rings have been replaced, the flange bolts should be re-tightened after
installation to compensate for seating of the O-rings.
4. If applicable, install flange adapters on the process end of the manifold using the 1.5-in. flange bolts
supplied with the transmitter.
Integral coplanar
Integral traditional
Traditional
Wafer
Summary of Contents for Rosemount MultiVariable 3051SMV
Page 2: ......
Page 6: ...vi Reference Manual 00809 0100 4853 Rev AB Contents January 2017 Contents ...
Page 44: ...36 Reference Manual 00809 0100 4853 Rev AB Installation January 2017 Installation ...
Page 62: ...54 Reference Manual 00809 0100 4853 Rev AB Troubleshooting January 2017 Troubleshooting ...
Page 111: ......