22
Reference Manual
00809-0100-4853, Rev AB
Installation
January 2017
Installation
Figure 3-4. Proper Bolt Installation
A. Bolt
B. Sensor module
O-rings with flange adapters
Table 3-1. Torque Values for the Flange and Flange Adapter Bolts
Bolt material
Head markings
Initial torque
Final torque
Carbon Steel (CS)
300 in-lb
650 in-lb
Stainless Steel (SST)
150 in-lb
300 in-lb
Failure to install proper flange adapter O-rings may cause process leaks, which can result in death or
serious injury. Only use the O-ring that is designed for its specific flange adapter.
A. Flange adapter
B. O-ring
C. PTFE based profile is square
D. Elastomer profile is round
Whenever the flange or adapters are removed, visually inspect the O-rings. Replace them if there are
any signs of damage, such as nicks or cuts. If the O-rings are replaced, re-torque the flange bolts and
alignment screws after installation to compensate for seating of the O-rings.
B7M
316
316
316
SW
316
STM
316
R
B8M
A
B
A
B
C
D