CAUTION
Reading errors
During operation, the debubbler is under pressure. A 0.040 in. (1 mm) orifice in the
outlet provides the pressure. Back pressure helps prevent outgassing, which can
lead to bubbles accumulating on the sensor face, resulting in erroneous readings.
Do not exceed 30 psig (308 kPa abs) inlet pressure.
The amount of pressure in the debubbler can be estimated from the flow rate. See
.
Table 2-2: Approximate Debubbler Pressure as a Function of Flow (0.040 Inch
Outlet Orifice)
gph
psig
mL/min
kPa abs
2
1
100
110
4
3
200
120
6
8
300
140
8
14
400
160
10
21
500
190
11
26
600
240
12
31
700
280
--
--
800
340
To control and monitor sample flow, a valved rotameter with fittings is available
(PN 24103-00).
3. Attach the rotameter to the debubbler outlet.
You can also use the rotameter to increase back pressure on the debubbler if
additional pressure is needed to prevent outgassing.
Install
Reference Manual
July 2020
00809-0100-3565
22
Emerson.com/Rosemount
Summary of Contents for Rosemount Clarity II T56
Page 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Page 4: ...4 ...
Page 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Page 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Page 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Page 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Page 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...