FM
approvals
Class I, Division 2, Groups A, B, C, and D
Class II, Division 2, Groups E, F, and G
T4 Tamb = -10 °C to 60 °C
1.2.2
Sensor
Method
ISO 7027 (using 860 nm LED source).
LED life
Three years (ISO 7027)
Wetted materials
Delrin
1
, glass, EPDM
Accuracy after
calibration at 20.0 NTU
0 - 1 NTU: ±2 percent of reading or ±0.015 NTU, whichever is
greater.
0 - 20 NTU: ±2 percent of reading
Note
The sensor can measure turbidity values of 2-200 NTU, but
frequent cleaning may be required to maintain turbidity
measurements.
Cable
20 ft. (6.1 m) or 50 ft (15.2 m). Connector is IP65.
Maximum pressure
30 psig (308 kPa abs)
Temperature
40 to 95 °F (5 to 35 °C)
Sensor body rating
IP65 when cable is connected.
1.2.3
Debubbler and flow chamber
Dimensions
18.1 x 4.1-in. (460 mm x 104 mm) diagram (approximately)
Wetted materials
ABS, EPDM, Delrin
®
, polypropylene, nylon
Inlet
Compression fitting accepts
¼
-in. OD tubing; fitting can be
removed to provide
¼
-in. FNPT.
Drain
Barbed fitting accepts
⅜
-in. ID tubing; fitting can be removed
to provide
¼
-in. female national pipe thread (FNPT). Must
drain to atmosphere.
Sample temperature
40 to 95 °F (5 to 35 °C)
Minimum inlet
pressure
3.5 psig (308 kPa abs).
3.5 psig will provide about 0.01 oz./min (250 mL/min) sample
flow.
Maximum inlet
pressure
30 psig (308 kPa abs). Do not block drain tube.
Recommended sample
flow
0.01 to 0.03 oz./min (250 to 750 mL/min)
Description and specifications
Reference Manual
July 2020
00809-0100-3565
10
Emerson.com/Rosemount
Summary of Contents for Rosemount Clarity II T56
Page 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Page 4: ...4 ...
Page 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Page 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Page 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Page 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Page 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...