4. If the material plugging the orifice cannot be removed with flushing, use a
toothpick or stiff wire to push out the obstruction. Push counter to the normal flow
through the orifice.
5. Reinstall the orifice and reconnect the drain line. Turn on the sample flow.
Postrequisites
If the blockage cannot be removed or the orifice is damaged during cleaning, replace the
orifice (PN 33947-00).
8.4
Replacement parts