Figure 5-3: Reset Screen
To reset settings, press
ENTER/MENU
from the main screen to access the
Reset
screen.
5.6
Program alarm relays
The Rosemount
™
T56 24 Vdc (-02 order code) and the AC switching power supply (-03
order code) provide four alarm relays for process measurement or temperature. You can
program each relay independently as an interval timer or one of four advanced timer
functions.
You can program the following relay functions to any relay from the
Relays
screen.
•
Assign a relay.
•
Define a relay function.
•
Assign a measurement.
•
Set relay logic.
•
Enter setpoints.
•
Set deadband.
•
Set normal state.
•
Set USP safety level (contacting conductivity).
Reference Manual
Program the transmitter
00809-0100-3565
July 2020
Rosemount T56
39