23
Reference Manual
00809-0100-4024, Rev CA
Mechanical Installation
March 2020
Mechanical Installation
Figure 2-5. Beam Width Distance, See
2.3.5
Special antennas and space requirements reference
Pipe installation
See
Installation requirements for the extended cone antenna
requirements for cone antenna in a still pipe/bridle
Extended cone installation
Installation requirements for the extended cone antenna
.
Table 2-4. Beam Width Distance
Antenna
Diameter of radiated area at different
distances from flange, ft. (m)
16 ft (5 m)
33 ft (10 m)
49 ft (15 m)
66 ft (20 m)
Cone 3 in.
7.2 (2.2)
14 (4.4)
22 (6.7)
29 (8.9)
Cone 4 in./ Process seal 4 in.
6.2 (1.9)
12 (3.7)
18 (5.6)
24 (7.4)
Cone 6 in./ Process seal 6 in.
5.2 (1.6)
10 (3.1)
15 (4.7)
21 (6.3)
Cone 8 in.
3.3 (1.0)
7.9 (2.4)
13 (3.9)
17 (5.2)
Parabolic
3.0 (0.9)
5.6 (1.7)
8.5 (2.6)
11 (3.5)
5m
10m
15m
20m
D
is
ta
nc
e
Summary of Contents for Rosemount 5600 Series