Section 8: programming Measurements
Instruction Manual
April 2017
LIQ-MAN-1066
48
Programming Measurements
Measure
The display screen for selecting the measurement is shown. The default
value is displayed in bold type. Refer to the contacting conductivity
Programming flow diagram to complete this function.
Range
The display screen for Selecting Auto-ranging or a specific range is
shown. The default value is displayed in bold type. Note: Ranges are
shown as conductance, not conductivity. Refer to the contacting
conductivity Programming flow diagram to complete this function.
Cell Constant
The display screen for entering a cell Constant for the sensor is shown.
The default value is displayed in bold type. Refer to the contacting
conductivity Programming flow diagram to complete this function.
RTD Offset
The display screen for Entering the RTD Offset for the sensor is shown.
The default value is displayed in bold type. Refer to the contacting
conductivity Programming flow diagram to complete this function.
RTD Slope
The display screen for entering the RTD slope for the sensor is shown.
The default value is displayed in bold type. Refer to the contacting
conductivity Programming flow diagram to complete this function.
Temp Comp
The display screen for Selecting Temperature Compensation as Slope,
Neutral Salt, Cation or Raw is shown. The default value is displayed in
bold type. Refer to the contacting conductivity Programming flow
diagram to complete this function.
8.4.3
8.4.4
8.4.5
8.4.6
8.4.7
8.4.8
1.234 µS/cm 25.0 ºC
S
N
Measurement
Conductivity
Resistivity
TDS
Salinity
NaOH (0-12%)
HCl (0-15%)
Low H2SO4
High H2SO4
NaCl (0-20%)
Custom Curve
1.234 µS/cm 25.0 ºC
SN
Range
Auto
50 µS
500 µS
2000 µS
20 mS
200 mS
600 mS
1.234 µS/cm 25.0 ºC
S
N
Cell Constant
1
.00000 /cm
1.234 µS/cm 25.0 ºC
S
N
RTD Offset
0
.00 °C
1.234µS/cm 25.0ºC
S
N
RTD Slope
2.00%/ºC
1.234 µS/cm 25.0 ºC
S
N
Temp Comp
Slope
Neutral Salt
Cation
Raw