Specifications
3
Instruction Manual
Section 2: Description and Specifications
LIQ-MAN-1066
April 2017
Features and applications
This loop-powered multi-parameter unit serves industrial, commercial and municipal applications
with the widest range of liquid measurement inputs available for a two-wire liquid transmitter.
The 1066 Smart transmitter supports continuous measurement of one liquid analytical input. The
design supports easy internal access and wiring connections.
analytical Inputs:
Ordering options for pH/ORP, Resistivity/Conductivity, % Concentration,
Total Chlorine, Free Chlorine, Monochloramine, Dissolved Oxygen, and Ozone.
large Display:
The high-contrast LCD provides live measurement readouts in large digits and shows
up to four additional variables or diagnostic parameters.
Digital Communications:
HART 7 and FOUNDATION Fieldbus options.
Menus:
Menu screens for calibrating and programming are simple and intuitive. Plain language
prompts and help screens guide the user through the procedures. All menu screens are available in
eight languages. Live process values are displayed during programming and calibration.
Quick Start programming:
Popular Quick Start screens appear the first time the unit is powered. The
instrument prompts the user to configure the sensor loop in a few quick steps for immediate commis-
sioning.
user help Screens:
Fault and warning messages include help screens similar to PlantWeb
™
alerts
that provide useful troubleshooting tips to the user. These on-screen instructions are intuitive and
easy to use.
Diagnostics:
The transmitter continuously monitors itself and the sensor for problems. A display
banner on the screen alerts Technicians to Fault and/or Warning conditions.
languages:
Emerson extends its worldwide reach by offering eight languages – English, French,
German, Italian, Spanish, Portuguese, Chinese and Russian.
Current Outputs:
HART units include two 4-20 mA electrically isolated current outputs giving the
ability to transmit the live measurement value and the process temperature reported from the sen-
sor.
Input Dampening:
is automatically enabled to suppress noisy process readings.
Smart-Enabled ph:
Rosemount SMART pH capability eliminates field calibration of pH probes
through automatic upload of calibration data and history.
automatic Temperature Compensation:
Most measurements require temperature compensa-
tion. The 1066 will automatically recognize Pt100, Pt1000 or 22k NTC RTDs built into the sensor.
Smart Wireless Thum adaptor Compatible:
Enable wireless transmissions of process variables
and diagnostics from hard-to-reach locations.
Section 2: Description and Specifications
2.1