HART Communications
113
1066Oz/Cl/DO
Instruction Manual
Section 10: haRT Communication
LIQ-MAN-1066
April 2017
1066Oz
1066Cl
Type
Enumeration / Range
Read/Write
Default
chlorine_type
The type of chlorine being measured.
ENUM
Free Chlorine
Total Chlorine
Chloramine
R / W
Free Chlorine
chlorine_resolution
The resolution of the displayed chlorine measurement.
ENUM
0.001 ppm
0.01 ppm
R / W
0.001 ppm
ph_correction_mode
Turns automatic pH compensation of free chlorine measure-
ments on and off.
ENUM
On
Off
R / W
Off
manual_ph
Provides a constant pH value, if manual pH compensation of a
free chlorine measurement is used.
FLOAT
0.00 -- 14.00 pH
R / W
7.00 pH
Type
Enumeration / Range
Read/Write
Default
chlorine_type
The resolution of the displayed ozone measurement.
ENUM
0.001
0.01
R / W
0.001
Type
Enumeration / Range
Read/Write
Default
polar_voltage
The polarization voltage used by the transmitter; automatically
set by sensor selection.
FLOAT
R / W
0
temp_coeff
The temperature coefficient used to compensate temperature;
automatically set by sensor selection.
FLOAT
R / W
0
amp_sensor_sensitivity
The response of the sensor to changes in concentration, deter-
mined by calibration.
FLOAT 0.1 -- 1,000,000.0
nA/ppm
R / W
2,500 nA/ppm
amp_sensor_zero_current
The current output of the sensor when the concentration is 0.
It is determined by a zero calibration.
FLOAT -999.9 -- 999.9 nA
R / W
0.0 nA