Instruction Manual
Section 8: programming Measurements
LIQ-MAN-1066
April 2017
Programming Measurements
43
Introduction
The 1066 automatically recognizes the measurement input upon first power-up and each time the
transmitter is powered. Completion of Quick Start screens upon first power up enable
measurements, but additional steps may be required to program the transmitter for the desired
measurement application. This section covers the following programming and configuration
functions:
1. Selecting measurement type or sensor type (all sections)
2. Identifying the preamp location (pH-see
3. Enabling manual temperature correction and entering a reference temperature (all sections)
4. Enabling sample temperature correction and entering temperature correction slope (selected
sections)
5. Defining measurement display resolution (pH and amperometric)
6. Defining measurement display units (all sections)
7. Adjusting the input filter to control display and output reading variability or noise (all sections)
8. Selecting a measurement range (conductivity – see
9. Entering a cell constant for a contacting or toroidal sensor (see
10. Creating an application-specific concentration curve (conductivity-see
11. Enabling automatic pH correction for free chlorine measurement (
To fully configure the transmitter, you may use the following:
1. Reset Transmitter function to reset factory defaults and configure to the desired
measurement. Follow the Reset Transmitter menu to reconfigure the transmitter to display
new measurements or measurement units.
2. Program menus to adjust any of the programmable configuration items. Use the following
configuration and programming guidelines for the applicable measurement.
8.1
Section 8: Programming Measurements