Instruction Manual
Section 8: programming Measurements
LIQ-MAN-1066
April 2017
Programming Measurements
55
Total Chlorine Measurement Programming
Description
This Chlorine sub-section describes how to configure the 1066 transmitter for Total Chlorine
measurement using amperometric chlorine sensors. The following programming and
configuration functions are covered:
1. Measure: Free Chlorine Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display resolution 0.01 or 0.001
4. Filter: 5sec Override the default input filter, enter 0-999 seconds
To configure chlorine measurement for total chlorine:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
The adjacent screen format will appear (factory defaults are shown). To program any displayed
function, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for chlorine programming at the end of Sec. 6 and
the 1066 live screen prompts for each function to complete configuration and programming.
Measure
The display screen for selecting the measurement is shown. The
default value is displayed in bold type. Refer to the chlorine
Programming flow diagram to complete this function.
Units
The display screen for selecting units as ppm or mg/L is shown. The
default value is displayed in bold type. Refer to the Chlorine
Programming flow diagram to complete this function
Filter
The display screen for entering the input filter value in seconds is
shown. The default value is displayed in bold type. Refer to the
Chlorine Programming flow diagram to complete this function.
8.6.2
8.6.2.1
8.6.2.2
8.6.2.3
8.6.2.4
1.234 µS/cm 25.0 ºC
S
N
Measurement
Free Chlorine
Total Chlorine
Monochloramine
1.234 µS/cm 25.0 ºC
S
N
Units
ppm
mg/L
1.234 µS/cm 25.0 ºC
S
N
Configure
Measure: Free Chlorine
Units: ppm
Filter: 5sec
Resolution: 0.001
1.234 µS/cm 25.0 ºC
S
N
Input filter
05 sec