Instruction Manual
Section 8: programming Measurements
LIQ-MAN-1066
April 2017
Programming Measurements
61
1.234µS/cm 25.0ºC
Program
Outputs
Measurement
T
emperature
Security
Reset
Analyzer
Frequency
1.234µS/cm 25.0ºC
Configure?
1.234µS/cm 25.0ºC
SN Configure
Measure: Cond
T
ype: 2-Electrode
Range:
Auto
Cell K: 1.00000/cm
R
T
D Of
fset: 0.00ºC
R
T
D Slope: 0
T
emp Comp: Slope
Slope: 2.00%/°C
Ref
T
emp: 25.0°C
Filter: 2 sec
Custom Setup
Measurement
contacting
1.234µS/cm 25.0ºC
SN
T
ype
2-Electrode
4-Electrode
1.234µS/cm 25.0ºC
SN Range
Auto
0-50 µS
50-500 µS
500-2000 µS
2000 µS - 20 mS
20 mS - 200 mS
200 mS - 600 mS
1.234µS/cm 25.0ºC
SN Cell Constant
1.00000 /cm
1.234µS/cm 25.0ºC
SN Measurement
Conductivity
Resistivity
TDS
Salinity
1.234µS/cm 25.0ºC
SN R
TD Of
fset
0.00°C
1.234µS/cm 25.0ºC
SN R
TD Slope
2.00%/ºC
1.234µS/cm 25.0ºC
SN
T
emp Comp
Slope
Neutral Salt
Cation
Raw
1.234µS/cm 25.0ºC
SN Slope
2.00 %/ºC
1.234µS/cm 25.0ºC
SN Ref
T
emp
(25.0ºC normal
)
+25.0ºC
1.234µS/cm 25.0ºC
SN Custom Curve
Configure
Enter Data Points
Calculate Curve
1.234µS/cm 25.0ºC
SN Custom Config
Units: ppm
# of Points: 2
Ref
T
emp: 25.0 ºC
Slope: 2.00 %/°C
1.234µS/cm 25.0ºC
SN Units
%
ppm
mg/L
g/L
1.234µS/cm 25.0ºC
SN Data Points
Pt1: 1.000 ppm
Pt1: 1.000 µS/cm
Pt2: 1.000 ppm
Pt2: 1.000 µS/cm
Pt3: 1.000 ppm
Pt3: 1.000 µS/cm
Pt4: 1.000 ppm
Pt4: 1.000 µS/cm
Pt5: 1.000
ppm
Pt5: 1.000 µS/cm
1.234µS/cm 25.0ºC
SN PointM
1.000 ppm
1.234µS/cm 25.0ºC
SN Calculate Curve
Custom curve
fit completed.
In Process Cal
recommended.
1.234µS/cm 25.0ºC
SN Calculate Curve
Curve fit in
progress …
Measurement
M
A
I
N
M
E
N
U
P
r
o
g
r
a
m
Figure 8-2. Configure Contacting Measurements