Instruction Manual
Section 8: programming Measurements
LIQ-MAN-1066
April 2017
Programming Measurements
57
Filter
The display screen for entering the input filter value in seconds is
shown. The default value is displayed in bold type. Refer to the
Chlorine Programming flow diagram to complete this function.
Resolution
The display screen for selecting display resolution as 0.001 or 0.01 is
shown. The default value is displayed in bold type. Refer to the
Chlorine Programming flow diagram to complete this function.
Oxygen Measurement programming
This section describes how to configure the 1066 transmitter for dissolved and gaseous oxygen
measurement using amperometric oxygen sensors. The following programming and
configuration functions are covered:
1. Sensor type: Select Water/Waste, Trace. BioRx, BioRx-Other, Brew, %O
2
In Gas
2. Measure type: Select Concentration, % Saturation, Partial Pressure, Oxygen in Gas
3. Units: ppm Select ppm, mg/L, ppb, g/L, % Sat, %O2-Gas, ppm Oxygen-Gas
4. Pressure Units: bar Select pressure units: mm Hg, in Hg, Atm, kPa, mbar, bar
5. Salinity: 00.0 %. Enter Salinity as %.
6. Filter: 5 sec Override the default input filter, enter 0-999 seconds
7. Partial Press: mmHg Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial pressure
To configure Oxygen:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
The adjacent screen format will appear (factory defaults are shown). To
program any displayed function, scroll to the desired item and press
ENTER.
The following sub-sections show the initial display screen that appears for each configuration
function. Use the flow diagram for oxygen programming at the end of Sec. 6 and the 1066 live
screen prompts for each function to complete configuration and programming.
8.6.3.3
8.6.3.4
8.7
1.234 µS/cm 25.0 ºC
S
N
Input filter
05 sec
1.234 µS/cm 25.0 ºC
S
N
Resolution
0.001
0.01
1.234µS/cm 25.0ºC
S
N
Configure
Type: Water/Waste
Units: ppm
Partial Press: mmHg
Salinity: 00.0‰
Filter: 5sec
Pressure Units: bar
Use Press: At Air Cal
Custom Setup