Serial Commands
P/N 400274-01 Revision A7
91
FF
- Follower Cycle Flags
The
Follower Cycle Flags
are a bit map in hexadecimal containing
selections for the control of operational and starting modes on a cycle by
cycle basis. More than one flag may be set or cleared at a time as the
flags are additive. It is on a individual cycle basis. The
Follower Cycle
Flags
are uploaded and downloaded by PCX.
Follower Defined By Sensor
0x01
Disable Cycle Counter Reset
0x08
Force Reset Master Processing
0x20
Disable Master Processing Reset Upon Fault
0x40
Following is an example of changing an operational flag:
CY = 5
Select the cycle five to edit.
FF?
Query the Follower flags on cycle 5 and PCX returns the
FF = 1
current flags for cycle 5:
The Follower cycle is defined by sensor.
The Master processing will be reset upon fault clearing.
Force Reset Master Processing
mode is disabled.
FF = 21
Adding 0x20 to the current value will enable
Force Reset
Master Processing
before starting on cycle 5.
Follower Cycle Defined By Sensor 0x01
This command specifies if this Follower cycle definition is defined by
sensor. It is available from PCX on the
Follower Cycles
screen.
Disable Cycle Counter Reset 0x08
This command allows the user to disable clearing the
Cycle Counter
when the Follower cycle is started. When active, the
Cycle Counter
will
be reset when the Master processing is reset. The default is disabled. See
KC4
serial command.
Force Reset Master Processing 0x20
This command allows the user to empty/reset the Master processing as if
a new
Master Cycle
definition was selected for the same Follower cycle.
This will occur every time the cycle is run. This causes the
Master Cycle
information to be loaded, thus overwriting the “learned” information
when a cycle starts.
Note:
This is only supported in Master defined by sensor mode.
The default setting for the
Force Reset Master Processing
flag is disabled
for all cycles.