Configuration
[50]
2.
The parameters
Point Id
,
Serial number
of the sensor, and
Serial number
of the cable on
tab
Input
.
3.
The parameter
Output sensitivity
on tab
Measurement
.
Major changes are:
1.
The parameters
Sensor
and
Extension cable
on the
Input
tab.
2.
All parameters on the
Measurement
tab except parameter
Output sensitivity
.
Figure 8-9: Button
Send & close
Note
Send & close
is disabled after major changes, which require a calibration of the converter, or if the
calibration state of the converter is not OK (see
Reload
Click
Reload
to reload the configuration including the calibration data from the converter to
the configuration editor. The data of the
Automatic calibration
and the
Offline calibration
(calibration process without Machine Studio) is loaded as
Manual calibration
data. Use the
report function to display the calibration data. Any configuration changes you made in
Machine Studio will be overwritten by the configuration loaded from the converter.
Figure 8-10: Button
Reload
Reset to factory default
Click
Reset to factory default
to reset the connected converter to the default parameter
settings.