Hazardous location installation
[34]
Table 6-4: Capacities and inductances of sensors including connection cable and
extension cable
(continued)
Sensor
Capacity (C
Sensor
)
Inductance (L
Sensor
)
EZ1900-003-ADAP-1 Adapter
cable for connection of PR 642x
sensors
+39 pF
negligible
EZ1900-003-ADAP-2 Adapter
cable for connection of BN
33010x sensors
+39 pF
negligible
Table 6-5: Temperature classification of sensors
EZ 105x and EZ 108x
T6 -35°C ≤ T
a
≤ 55°C
T4 -35°C ≤ T
a
≤ 130°C
T2 -35°C ≤ T
a
≤ 200°C
PR 6422
T6 -35°C ≤ T
a
≤ 60°C
T4 -35°C ≤ T
a
≤ 130°C
T2 -35°C ≤ T
a
≤ 150°C
PR 6423
T6 -35°C ≤ T
a
≤ 65°C
T4 -35°C ≤ T
a
≤ 130°C
T2 -35°C ≤ T
a
≤ 200°C
PR 6424
T6 -35°C ≤ T
a
≤ 65°C
T4 -35°C ≤ T
a
≤ 130°C
T2 -35°C ≤ T
a
≤ 155°C
PR 6425
T6 -35°C ≤ T
a
≤ 65°C
T4 -35°C ≤ T
a
≤ 130°C
T2 -35°C ≤ T
a
≤ 275°C
PR 6426
T6 -35°C ≤ T
a
≤ 65°C
T4 -35°C ≤ T
a
≤ 130°C
T2 -35°C ≤ T
a
≤ 155°C
PR 642x high temperature
types
T6 -35°C ≤ T
a
≤ 65°C
T4 -35°C ≤ T
a
≤ 130°C
T2 -35°C ≤ T
a
≤ 200°C
Summary of Contents for AMS EZ1000