Technical Specifications
11.1 Ambient Conditions
The SonoTT
TM
Ultrasonic FlowComputer and the SonoTT
TM
Clamp-On Transducer are intended to be used
under the following ambient conditions:
Operating altitude
max. 3000m
Air pressure
70 to 106 kPa
Operating temperature range
10° C to 40°C (50°F to 104°F)
Storage temperature range
–20°C to 45°C (- 4°F to 113°F)
Transport temperature range
–20°C to 55°C (- 4°F to 131°F)
Relative Humidity:
Operation
Storage and transport
30 % to 75 % (non-condensing)
10 % to 96 % (non-condensing)
11.2 SonoTT
TM
Ultrasonic FlowComputer
Size (HxWxD)
90 x 210 x 293 mm
Weight
max. 4 kg
Case material
stainless steel
Protection Class
Class I equipment
IP-Code
IPX1 (Protected against vertically falling water drops)
Supply voltage
100-240VAC / 50-60Hz
Power consumption (without printer)
max. 60 VA
Fuses
250V; 1.6AL; Type T; 5x20 mm
Mains Power Cord EU
type EU, l = 2m
Interfaces:
Interface type
RS-232
Device connector
9-pin D-Sub-socket
RS-232 cable
shielded, 9-pin D-Sub-cable, (Male/Female), l=2 m
Connection limitations
connect only devices tested according to
IEC 60601
or
IEC 60950 2
nd
ed/IEC 62368 1
st
ed.
Maximum flow measurement range
± 32,0 liter per minute (lpm)
(depending on calibration and type of used SonoTT
TM
Clamp-On
Transducer)
Resolution of flow measurement values:
0 to 999 mlpm
1 milliliter per minute (mlpm)
1.00 to 9.99 lpm
10 mlpm
10.0 to 32.0 lpm
100 mlpm
Information Alarms