Changing the Setup Parameters
The device and the parameters of the SonoTT
TM
Clamp-On Transducer can be changed as follows:
•
Activation of the desired field.
•
Changing the value by rotating of the rotary switch.
•
Confirming of the new value by pressing the integrated push-button.
•
If desired, likewise change the values in the other fields.
•
Leave the
Settings
menu and save the settings by activating the
Exit/Save
field .
•
The values are then taken over by the system and remain valid with the next power-up sequence.
If the values don't change, exit the menu by activating the Exit field.
6.5 Activation of the Numeric Display Mode (Numeric)
In the
Numeric
menu, the mean value of a single measurement system (averaged over 3 seconds) is numerically
displayed in the main display field. In the monitor display field, all activated measurement systems are
displayed:
Q1 = Flow Measuring System 1
Q2 = Flow Measuring System 2
Each measurement system in the monitor display field can be activated in the main field by selecting and
activating the desired abbreviation (Q1, Q2) in the monitor display field. The selected measurement system
has a purple background (e.g.
Q1
).
The measured flow values (
Q1
and/or
Q2
) are shown in litres per minute (lpm). The upper limit of the flow
value is dependent upon the transducer but is limited to ± 32.0 lpm. In the range from ± 1.00 to 9.99 lpm there
are two digits after the decimal and above ±10.0 lpm one digit after the decimal. Under ±1 lpm the measured
flow is displayed in mlpm.