EN
10 Record of Safety Test for SonoTT
TM
Ultrasonic FlowComputer and Its
Accessories*
*accessories according these operating instructions
General
Test organisation,
Name of Tester
Initial test (reference value):
Recurrent test:
Test after repair:
Device
SonoTT
TM
Ultrasonic FlowComputer
Type
Reference no.
SN
Manuf. year
Protection class
I
Applied parts (Type, SN)
Accessories
Note: The following safety checks should be performed at least every 24 months by a qualified person who has adequate
training, knowledge, and practical experience to perform these tests and be subject to the inspection activity to any
directives.
Visual Inspection
Visual check for mechanical damage that could impair its functions.
Visual inspection
Passed?
Device & accessories configured according this operating instructions and not changed?
Yes
No
Power supply cord
Yes
No
Other connection cables
Yes
No
Housing
Yes
No
Soiling
Yes
No
Is the type plate legible?
Yes
No
Are the safety relevant markings legible?
Yes
No
Are the accessories used compatible?
Yes
No
Electrical Safety
The device is tested at the factory prior to delivery, please use the values listed in the accompanying
inspection record as targets values. Perform the inspection in accordance with
IEC/EN 62353 2
nd
ed.
. If the
inspection reveals significant deviations, contact our Service department. We recommend the safety tester
manufactured by GERB Elektronik as the testing instrument.
Testing the electrical safety
Limit value
Target value
Actual value
Passed?
Measurement of power consumption
< 60 VA
without printer
Yes
No
SonoTT
TM
Ultrasonic FlowComputer
Page 84
D108-702 SonoTT
TM
Ultrasonic FlowComputer - Operating Instructions V4.0
Copyright em-tec GmbH