67
A
pe
Vi
PAT1
PAT2
FE
Leakage Current Tester 92-4A, -4D, -4E, -4G
MEASURING CIRCUITS
II.ii
Measuring Circuits (Connection principle)
The following circuits show the connection principle of the leakage cur-
rent measurement for various variants. The actual connection depends
on the item under test itself or the measurement instrument. Details on
the possible connection wirings and regulations contain the respective,
for the item under test valid standards.
Enclosure
Illustr. II.6: Connection
principle for Ground
Leakage Current
GLC
E.g. of a motor housing pe
Protective insulation Vi
Illustr. II.7: Connection
principle Enclosure
Leakage Current
ELC
Illustr. II.8: Connection principle for Patient Leakage Current
PLC
and Patient Auxiliary Current
PAC