80
A
Leakage Current Tester 92-4A, -4D, -4E, -4G
BASIC TEST MODES
VI.vii Measuring of contact currents according EN 60598
The measuring of the contact current is carried out according the network
shown below. The contact current have to observe the values predefined
in the EN-standard. A detailed description of the prerequisites for the
measuring contains the EN-standard (see above).
Measurement circuit according EN 60990-1 Fig. 3/4 or 3/5.
Circuit diagram:
Class I to PE
A1 - normal polarity without fault
A2 - inverse polarity without fault
A3 - normal polarity with N-fault
A4 - inverse polarity with N-fault
Class I to Vi
A1 - normal polarity without fault
A2 - inverse polarity without fault
A3 - normal polarity with PE-fault
A4 - inverse polarity with PE-fault
A5 - normal polarity with N-fault
A6 - inverse polarity with N-fault
Class II
A1 - normal polarity without fault
A2 - inverse polarity without fault
A3 - normal polarity with N-fault
A4 - inverse polarity with N-fault
Illustr. VI.7: Basic measuring network for leakage current
measurement for Class I and II test objects.