Getting Started
90
Automatic Priming Using the Pump
Before commencing infusion setup, the administration set needs to be primed. Priming
expels all the air from the administration set, and fills it with the infusion liquid. A fully
primed administration set is a set filled with infusion liquid (from which all the air was
removed).
Priming with the pump can be initiated from the following screens:
•
Start Up
•
Start
•
Paused (infusion or bolus)
•
Air in Line Alarm
Before using the pump for priming, ensure that:
•
The administration set clamp is open.
•
The safety door is closed.
•
The Q Core administration cassette is properly connected to the pump.
It is recommended to prime the administration set using the pump.
Before priming, verify that the administration set is disconnected
from the patient.
When the Prime Reminder is enabled and the set has not yet been
primed using the pump, a Prime Reminder will be displayed,
enabling the user to press
Prime
in order to start priming or to
press
Start
in order to proceed with the infusion (refer to
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......