Using the Delivery Modes
150
Infusion Parameters: PCA Mode
The following infusion parameters need to be set for a PCA infusion:
Parameter
Description/Notes
Continuous Rate
The rate of the basal infusion. Continuous Rate values can range
from 0.1 to 99.9 mL/h, or be equal to zero (Bolus Only infusion).
VTBI
The total amount of fluid to be infused. The remaining VTBI is
displayed on the screen as the infusion progresses. VTBI values can
range from 0.1 to 9999 mL.
Demand Bolus
The amount of fluid infused in a single bolus. Demand bolus values
can range from 0.1 to 30 mL, or be equal to zero (Continuous Only
infusion).
Bolus Lockout
The minimum time that must pass between the end of one bolus to
the start of the next bolus. After a bolus delivery ends, the next
bolus becomes available following the lockout time.
Dose Limit
The option to choose if patient boluses are limited by number or
by volume. When choosing No Limits, the patient boluses are set
to the maximum allowed volume, according to the other
parameters defined for the infusion, including lockout time and
demand bolus.
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......