Technical Specifications
270
The following sections present technical specifications for the pump and its
components:
....................................................................................................
.............................................................................................
Average Bolus Volume After Occlusion
..................................................................
................................................................................
Electromagnetic Compatibility Statement
..............................................................
Pump Accuracy
The following graphs and curves were derived from the pump accuracy testing
procedures described in the IEC60601-2-24 standard. Testing was performed under
normal conditions (specified in IEC60601-2-24 standard) at room temperature (25°C,
72°F).
Normal conditions to ensure optimal accuracy of ±2.5%:
•
Fluid level should be 50 cm above the pump
•
No back pressure due to catheter size or difference in height of pump and infusion
site
•
Room temperature (25°C)
•
Barometric pressure of sea level altitude (101kPa)
•
IV medication with water like fluid characteristics
Chapter 12: Technical Specifications