Sapphire Infusion Pump User Manual
281
Electromagnetic Compatibility Statement
The following sections provide information about testing of and recommendations for:
.....................................................................................
...................................................................................
Recommended Separation Distances from Mobile RF Communication Equipment
Electromagnetic Emission
The pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the pump should ensure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment
Guidance
RF emission CISPR 11
Group 1
The pump uses RF energy for its internal
function only. Therefore, RF emissions are
very low and are not likely to cause any
interference with nearby electronic
equipment.
RF emission CISPR 11
Class B
The pump is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 610003-3-2
Class B
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
Complies
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......