Sapphire Infusion Pump User Manual
199
Setting Delivery Mode
The Sapphire pump is a multi-platform device that has the ability to operate in several
different modes. Each delivery mode features its own unique options. The current
mode is displayed at the right side of the Indicators Bar.
Setting the delivery mode is done using the
Options
menu. An authorization code of
High or Technician is required to modify the delivery mode.
> To c h a n g e t h e d e l i v e r y m o d e :
1.
From the toolbar, press
Options
.
2.
From the
Options
screen, select
Delivery mode
.
3.
Using the keypad, enter the required password, and then press
OK
.
4.
Select the relevant delivery mode.
The
Start Up
screen of the selected mode appears.
When you change the delivery mode, the pump reverts to the
default values for the newly selected mode.
To access the PCEA or Intermittent Epidural delivery modes, select
Epidural
. Then select the relevant Epidural delivery mode.
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......