Technical Specifications
284
Notes
1.
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz
are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to
27,283 MHz; and 40,66 MHz to 40,70 MHz.
2.
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz
and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the
likelihood that mobile/portable communications equipment could cause
interference if it is inadvertently brought into patient areas.
For this reason, an additional factor of 10/3 is used in calculating the
recommended separation distance for transmitters in these frequency ranges.
where
p
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer, and
d
is the
recommended separation distance in meters
(m).
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2,5 GHz
10 V/m
80 MHz – 2.5 GHz
80% AM, 2 Hz
80 MHz – 2.5 GHz
80% AM, 1 KHz
Recommended
separation distances:
•
80 MHz to 800 MHz:
•
800 MHz to 2.5 GHz:
At 80 MHz and 800 MHz, the higher frequency
range applies.
* These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
† The pump was tested to the EMC requirements of
IEC 60601-1-2 (third edition)
and
IEC 60601-2-24 (second edition)
.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment
Guidance
*
d
4
p
=
d
7.7
p
=
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......