Options Menu: Configuring, Viewing and Testing
206
Advanced Bolus
Allows users to program a bolus by
entering rate, amount and time. When this
option is disabled, the bolus is
programmed by amount only, and the rate
is the default bolus rate. The option is
available only when Allow Bolus is enabled
(by a Technician authorization code).
Applicable only for the Continuous
delivery mode.
For more information, refer to
.
Select the
Advanced
Bolus
row, to toggle the
option between
On
and
Off
.
Bolus Reminder
Enables a reminder for the user to connect
the bolus handle before starting a PCA,
PCEA or PIEB infusion that includes patient
boluses. The reminder:
•
Instructs to connect the bolus handle
directly to the pump.
•
Checks functionality – bolus press is
recognized by the pump.
Select the
Bolus
Reminder
row, to toggle
the option between
On
and
Off
.
Auto P. lockout
Enables/disables Patient Lockout, a safety
feature that requires password entry to
make any parameter changes.
When the option is enabled, Patient
Lockout is activated automatically as the
infusion begins (
Select the
Auto P.
Lockout
row, to toggle
the option between
On
and
Off
.
Option
Descriptions/Notes
To Modify Parameter
(from the General
settings screen)
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......