Components, Accessories, and Administration Sets
50
•
Output voltage: 10 VDC
> To c o n n e c t t h e A C P o w e r C o rd t o t h e I P S :
1.
Connect the Q Core AC Power Cord to the IPS.
2.
Connect the Cord Retainer to the IPS, using a Philips head screwdriver.
> To c h a rg e t h e b a t t e r y :
1.
Plug the Q Core AC Power Cord into the main power supply source.
2.
On the back of the IPS, verify that the Power LED status indicator is On (blue light).
3.
Mount the pump into the mini cradle.
4.
On the front of the pump, verify that the Charge LED status indicator is On
(blinking yellow light).
> To d i s c o n n e c t t h e A C P o w e r C o rd f ro m t h e I P S :
1.
Unplug the Q Core AC Power Cord from the main power supply source.
2.
Disconnect the Cord Retainer from the IPS, using a Philips head screwdriver.
3.
Remove the Q Core AC Power Cord from the IPS.
The cord retainer is an integral part of the IPS, and must be
connected prior to initial use.
Summary of Contents for Q Core Sapphire
Page 1: ......
Page 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Page 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Page 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Page 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Page 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Page 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Page 288: ...Sapphire Infusion Pump User Manual 287 ...
Page 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Page 293: ......
Page 294: ......