Electrical Adjustment
Grp / No. Item
Function
Range
Initial
Note
241
Flow Ideal Data
(Value will be set after calibration)
0 Flow Ideal Data 1
Flow Sensor Theory Value No.1 Corresponding to Fan Volt No.1
0 - 1023
330
1 Flow Ideal Data 2
Flow Sensor Theory Value No.2 Corresponding to Fan Volt No.2
0 - 1023
390
2 Flow Ideal Data 3
Flow Sensor Theory Value No.3 Corresponding to Fan Volt No.3
0 - 1023
420
3 Flow Ideal Data 4
Flow Sensor Theory Value No.4 Corresponding to Fan Volt No.4
0 - 1023
435
4 Flow Ideal Data 5
Flow Sensor Theory Value No.5 Corresponding to Fan Volt No.5
0 - 1023
440
5 Flow Ideal Data 6
Flow Sensor Theory Value No.6 Corresponding to Fan Volt No.6
0 - 1023
1023
6 Flow Ideal Data 7
Flow Sensor Theory Value No.7 Corresponding to Fan Volt No.7
0 - 1023
1023
7 Flow Ideal Data 8
Flow Sensor Theory Value No.8 Corresponding to Fan Volt No.8
0 - 1023
1023
242
Flow Calib. Fan Volt
(Not Memorized)
0 Flow Calib Fan Volt 1
Flow Sensor Calibration Fan Volt No.1 [V x10]
0 - 255
32
1 Flow Calib Fan Volt 2
Flow Sensor Calibration Fan Volt No.2 [V x10]
0 - 255
46
2 Flow Calib Fan Volt 3
Flow Sensor Calibration Fan Volt No.3 [V x10]
0 - 255
61
3 Flow Calib Fan Volt 4
Flow Sensor Calibration Fan Volt No.4 [V x10]
0 - 255
84
4 Flow Calib Fan Volt 5
Flow Sensor Calibration Fan Volt No.5 [V x10]
0 - 255
100
5 Flow Calib Fan Volt 6
Flow Sensor Calibration Fan Volt No.6 [V x10]
0 - 255
255
6 Flow Calib Fan Volt 7
Flow Sensor Calibration Fan Volt No.7 [V x10]
0 - 255
255
7 Flow Calib Fan Volt 8
Flow Sensor Calibration Fan Volt No.8 [V x10]
0 - 255
255
243
Flow Calib Effective Difference
(Not Memorized)
0 Flow Calib Effective Diff 1
Flow Sensor Calibration Range Diff No.1 [ADC]
0 - 1023
180
1 Flow Calib Effective Diff 2
Flow Sensor Calibration Range Diff No.2 [ADC]
0 - 1023
190
2 Flow Calib Effective Diff 3
Flow Sensor Calibration Range Diff No.3 [ADC]
0 - 1023
200
3 Flow Calib Effective Diff 4
Flow Sensor Calibration Range Diff No.4 [ADC]
0 - 1023
210
4 Flow Calib Effective Diff 5
Flow Sensor Calibration Range Diff No.5 [ADC]
0 - 1023
215
5 Flow Calib Effective Diff 6
Flow Sensor Calibration Range Diff No.6 [ADC]
0 - 1023
1023
6 Flow Calib Effective Diff 7
Flow Sensor Calibration Range Diff No.7 [ADC]
0 - 1023
1023
7 Flow Calib Effective Diff 8
Flow Sensor Calibration Range Diff No.8 [ADC]
0 - 1023
1023
244
Flow Clog Threshold
(Not Memorized )
0 Flow Clog Threshold 1
Flow Sensor Clogged Threshold No.1[ADC]
0 - 1023
260
1 Flow Clog Threshold 2
Flow Sensor Clogged Threshold No.2[ADC]
0 - 1023
320
2 Flow Clog Threshold 3
Flow Sensor Clogged Threshold No.3[ADC]
0 - 1023
350
3 Flow Clog Threshold 4
Flow Sensor Clogged Threshold No.4[ADC]
0 - 1023
400
4 Flow Clog Threshold 5
Flow Sensor Clogged Threshold No.5[ADC]
0 - 1023
430
5 Flow Clog Threshold 6
Flow Sensor Clogged Threshold No.6[ADC]
0 - 1023
1023
6 Flow Clog Threshold 7
Flow Sensor Clogged Threshold No.7[ADC]
0 - 1023
1023
7 Flow Clog Threshold 8
Flow Sensor Clogged Threshold No.8[ADC]
0 - 1023
1023
10 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
360
11 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
420
12 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
450
13 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
500
14 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
530
15 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
1023
16 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
1023
17 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
1023
246
Flow Clog Monitor
(Read only)
0 Real Flow Data
Real Flow Data [ADC] (Read only)
0 - 1023
-
1 Ideal Flow Data
Theory Flow Data (Read only)
0 - 1023
-
2 Flow Difference Data
Flow Difference Data [ADC] (Read only)
-1023 - 1023
-
3 Clog Detect Flow Difference Data Difference Data to judge the Filter Clogged [ADC] (Read only)
0 - 1023
-
4 Warn Detect Flow Difference Data Difference Data to judge the Filter Clogged Warning [ADC] (Read
only)
0 - 1023
-
5 Press Add Data
Add Data by Air Pressure (Read only)
0 - 1023
-
6 Filter Scroll Timer
Timer for Filter Scroll [hour] (Read only)
0 - 32767
-
7 Air Inlet Err Difference Data
Difference to judge the Intake Error
0 - 1023
-
249
Pressure Sensor
0 Pressure Monitor
Pressure Sensor Value [ADC] (Read only)
-
-
1 Pressure Monitor
Pressure Sensor Value with Compensate [ADC] (Read only)
-
-
2 Presss Offset
Offset Value
- 1023 - 1023
0
3 Press calib Effective Diff
Effective Range [ADC]
0 -123
1023
4 Press Sensor Calibration Log
0: Executed & Normal Range (Read ony)
10: Out of Range Minimum
20: Out of Range Maximum
99: Not execute Calibration
-
99
252
Fan Option
1 Safety Switch
Safety Application use
0 - 8
0
2 Fan Manual Switch
Fan Manual SW (0: Auto 1: Manual)
0 - 1
0
Summary of Contents for LC-XT6
Page 192: ...192 IC Block Diagrams CXD3548 Gamma IC401 CXA7009 S H IC501 IC531 IC561 IC1501 IC1531 IC1561...
Page 193: ...193 IC Block Diagrams FA5501 PFC IC1601 IC1651 HIN202 RS232C Driver IC3801...
Page 195: ...195 IC Block Diagrams AX11005 Network IC8301 TE7783 I O Expander IC1801...
Page 196: ...196 IC Block Diagrams...
Page 204: ...SPL 8 LC XT6 Exploded Views M01 4 M01 5 Lens shift assembly M01 1 M01 3 M01 2...
Page 209: ...SPL 13 LC XT6 Exploded Views Optical filter LC CS L19 Integrator assembly S06 L11 S06...
Page 211: ...SPL 15 LC XT6 Exploded Views Relay lens OUT assembly L06 S06 S06...
Page 212: ...SPL 16 LC XT6 Exploded Views L08 L07 In the Optical unit L15 L15 L05 L22 L14 L13 L21 L09 L10...
Page 213: ...SPL 17 LC XT6 Exploded Views Optical filters assembly Mirror assembly L16 L16 L18 L17...
Page 214: ...SPL 18 LC XT6 Exploded Views Labels W09 W07 W01 W01 W06 W08 W04 W04 W04 W03 W02 W05...
Page 218: ...SPL 22 LC XT6 Mechanical Pats List...