Manual Cleaning Instruction
1. Remove debris
Rinse the device under running cold tap water for a minimum of 2 minutes. Use a sponge, soft lint-free cloth or soft-
bristled brush to assist in removing gross soil. For cannulations of the handpiece and attachments, the cleaning brush
(item no. 30611807) should be used.
Figure 23
Precautions
• Never immerse the handpiece, lid or attachments in aqueous solutions or in an ultrasonic bath as this could decrease
the service life of the system.
• Do not use pointed objects for cleaning.
2. Manipulate Moving Parts
Manipulate all moving parts such as the triggers, sleeves and switches under running tap water to loosen and remove
gross debris.
3. Spray and Wipe
Spray and wipe the device using a neutral pH enzymatic solution for a minimum of 2 minutes.
Follow the enzymatic detergent manufacturer’s directions for correct temperature, water quality (i. e. pH, hardness) and
concentration / dilution.
4. Clean with Detergent
Clean the device manually under running warm water using an enzymatic cleaner or detergent for a minimum of 5 minutes.
Manipulate all moving parts under running water. Use a soft-bristled brush and / or soft lint-free cloth to remove all visible
soil and debris. Follow the enzymatic cleaner or detergent manufacturer’s instructions for use for correct temperature,
water quality and concentration / dilution.
5. Rinse with Tap Water
Rinse the device thoroughly using cool to lukewarm running water for a minimum of 2 minutes. Use a syringe, pipette or
water jet to fl ush lumens and channels. Actuate joints, handles and other movable device features in order to rinse
thoroughly under running water.
6. Visually Inspect Device
Inspect the cannulations, coupling sleeves, etc. for visible soil. Repeat steps 1 – 6 until no visible soil remains.
7.
Final Rinse with De-Ionized / Purifi ed Water
Final rinse with de-ionized or purifi ed water for a minimum of 2 minutes.
8. Dry
Dry device using a soft lint-free cloth or clean compressed air.