12
USER MANUAL
ORTHOVET PLUS BATTERY DRILL SYSTEM
T +49 7461 96 580 0 | F + 49 7461 96 580 90 | [email protected] | www.eickemeyer.com
Quick Chuck for KIRSCHNER-Wires (Item No. 195079)
KIRSCHNER-Wires of any length with a diameter of 0.6 – 2.6 mm can be used with the Quick Chuck for KIRSCHNER-Wires.
1. Adjust the KIRSCHNER-Wire diameter according to the label on the adjusting sleeve • •• ••• (Fig. 16, no. 1). Slightly press
the adjusting sleeve axially against the handpiece and rotate the sleeve.
2. Apply a slight amount of pressure to insert the KIRSCHNER-Wire from the front into the cannulation.
The wire is held automatically.
3. Adjust the working length by pulling on the wire.
4.
To affi x the wire, pull the tension lever (Fig. 16, Nr. 3) against the tool with your little fi nger and ring fi nger. Only pull the
tension lever against the tool as much as necessary. The clamping force can be varied by pulling and releasing the
clamping lever.
5. Insert the wire into the bone. Apply the clamping force as long as the wire is advanced.
6. To adjust the grip on the wire, reduce the clamping force and move the tool to the desired length. Reclamp the wire by
pulling on the tension lever.
2
3
1
Figure 16
Oscillating Saw-Chuck (Item No. 195081)
For saw blades with Aesculap-Connection.
Figure 17