7. Load Tray / Washing Basket
Place devices in the specially designed tray for machine washing supplied by EICKEMEYER® (item no. 195086).
Dimensions load tray / washing basket (in mm): L 260 x W 170 x H 110
Ensure that all cannulations (handpiece and attachments), if applicable, are positioned as shown (Figure 25 + 26).
Figure 25
Figure 26
Note:
For the tray / washing basket a corresponding container (item no. 185555, Fig. 27).
• Non perforated bottom, perforated lid
• Dimensions
bottom
(in mm): L 312 x W 183 x H 122
Figure 27
8. Automated Cleaning
Step
Duration (minimum)
Cleaning instructions
Rinse
2 minutes
Cold tap water
Pre-wash
1 minute
Warm water (≥ 40 °C), use detergent
Cleaning
2 minutes
Warm water (≥ 45 °C), use detergent
Rinse
5 minutes
Rinse with de-ionized (DI) or purifi ed water (PURW)
Thermal disinfection
5 minutes
Hot de-ionized (DI) water, ≥ 93 °C
Dry
40 minutes
≥ 90 °C
9. Inspect the Device
Remove all the devices from the tray / washing basket. Inspect the cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual pre-clean / automated cleaning cycle. Confi rm that all parts are completely dry. Mechanical
cleaning / disinfection is an additional stress for power equipment, especially for seals and bearings. Therefore, systems
must be properly lubricated and regularly sent to be serviced (at least once per year).