5
USER MANUAL
ORTHOVET PLUS BATTERY DRILL SYSTEM
T +49 7461 96 580 0 | F + 49 7461 96 580 90 | [email protected] | www.eickemeyer.com
Effi ciently working cutting tools are the basis for successful surgery. Therefore, it is mandatory to check used cutting tools
after every use for wear and / or damage and to replace them if necessary. We recommend using new EICKEMEYER® cutting
tools for every surgery. Cutting tools must be cooled with irrigation liquid to prevent heat necrosis. The user of the product is
responsible for proper use of the equipment during surgery. If the OrthoVet Plus System is used in conjunction with an
implant system, make sure to consult the corresponding “Technique Guide”.
The manufacturer shall assume no responsibility for damage resulting from improper operation, neglected or unauthorized
maintenance of the tool.
Precautions
• To avoid injuries, the locking mechanism of the tool has to be activated before every manipulation and before placing
the tool back down, i. e. the mode switch has to be in “OFF” position.
• The handpiece may only be operated with a fully charged battery. To do this, ensure that the battery is charged in good time.
EICKEMEYER® recommend that the battery is replaced into the charger immediately after surgery.
• The batteries may never be sterilized, washed or dipped. Sterilization would destroy the battery with possible secondary
damage.
•
Should the machine drop on the fl oor and have visible defects, do not use it anymore and send it to the EICKEMEYER®
service center.
Accessories / Scope of Delivery
The OrthoVet Plus Battery Drill System consists of a handpiece, one battery loading station two batteries and a range of
attachments and accessories designed for the system.
Special auxiliaries such as cleaning brushes and oil are available for cleaning and servicing the system. No oils from other
manufacturers must be used. Only lubricate the power tool and the attachments when clean.
EICKEMEYER® recommends the use of the specifi cally designed EICKEMEYER® OrthoVet Plus Tray (item no. 195986) and
Container-System (item no. 185555) to sterilize and store the system.
The following components are essential to ensure proper operation:
• Handpiece (item no. 195077)
• Battery charging station (item no. 195082)
• Battery (item no. 195083)
• Funnel for handpiece (item no. 195085)
• At least one attachment / chuck of the system (item no. 195078, 195079, 195080 or 195081)
Please refer to the end of these Instructions for Use for an overview of the components of the system.
Storage and Transport
Please use the original packaging / show case (item no. 195088) for dispatch and transport. If this is no longer available,
please contact your EICKEMEYER® offi ce.
Warranty
The warranty for the tools and accessories does not cover damage of any kind resulting from improper use, damaged seals or
improper storage and transport. EICKEMEYER® does not accept liability for damage resulting from repairs or maintenance
carried out by unauthorized sites.