8-71
CAUTION!
EdgeTech recommended performing the process described in the section
below through Discover, but the process below provides an alternative.
to complete this process in Discover.
8.3.5
Applanix POSMV Configuration
The 6205s is compatible with the POSMV system from Applanix. Follow the below steps to configure the
6205s and the POSMV for communication over the Ethernet interface via UDP.
Utilizing the Applanix POSVIEW software:
1.
Set the ethernet real-time output data to groups 3,7,10, and 102. The output rate of the POSMV
data should be set to 50Hz.
2.
The POSMV must also be set to Falling (Negative) Edge or
Active Low
for the 1PPS as the 6205s
defaults to sync on the falling edge.
3.
Set the POSMV Time Tag 1 to UTC.
To configure the 6205s COM ports for the UDP messages from the POSMV:
1.
2.
Under [Serial1], [Serial2], and [Serial3] set Create = 0.
3.
Under [Serial5], set Create = 1 and remove the semicolon from the beginning of each line in the
[Serial5] section.
4.
Ensure the command Parserparameters=HEAVESCALE -1.0 is present in the [Serial5] section. If it
is not present, add it.