System Overview
| 7
14.
Connect the Eddyfi supplied subsea whip to the subsea bottle umbilical connector
15.
Connect the Eddyfi supplied topside whip free end relevant communication wires to the
ROV topside integration port. See Connectors on page 24. Connect the other end to
the topside unit umbilical connector.
16.
Connect the probe(s) to be used to the U41 subsea unit. Ensure that any probe connector
which doesn’t have a probe connected to it is fitted with a blanking plug. Deploying the
bottle subsea with an exposed connector will cause severe damage to the unit.
17.
Power up the topside unit and ROV port. Using ASSIST, attempt to connect to the subsea
bottle. If there is no connection, re-check the physical connections and, if using serial
protocols, try swapping Rx and Tx at one end of the ROV interface.
18.
Once communications are established, function check the probe on the test plate to
ensure that data is passing reliably through the system.