System Overview
| 13
Figure 1-18
Operation with ROV - Subsea bottle configuration
A dialogue similar to the one shown below should appear.
Figure 1-19
Operation with ROV - Subsea bottle configuration
If a standard ethernet connection is available through the ROV, untick Enable Serial and set
Adaptor to Ethernet.
If using serial communications direct to the U41 bottle or through a ROV multiplexing system,
tick Enable Serial, then specify protocol and baud rate using Mode and Speed (bps).
If using Eddyfi supplied subsea umbilical cable direct from Topside Unit to U41 subsea bottle, set
Adaptor to Ext. Ethernet.
Note: If you intend to connect directly to the ASSIST PC, i.e. not via the Topside Unit, using
standard Ethernet, then select DHCP Server as the IP Mode as the last configuration change
here.
To apply the changes, press the Green Arrow button. This will close the dialogue, restart the U41
bottle and show the Instrument Connection dialogue once again. There will be a delay before
the U41 instrument reappears in the dialogue.
If the Topside Unit has already been configured with matching serial protocols, after another
delay, a second ‘U41 – Serial’ instrument connection method will appear similar to that shown
in the dialogue below.
Summary of Contents for TSC U41
Page 1: ...Getting Started with TSC U41 Modernizing Subsea ACFM User s Manual ...
Page 7: ...General Precautions and Conventions ...
Page 13: ...Chapter 1 System Overview ...
Page 27: ...Chapter 2 The Subsea Bottle ...
Page 33: ...Chapter 3 The Topside Unit ...
Page 40: ...Chapter 4 Assist3 Software ...
Page 67: ...Assist3 Software 55 Figure 4 37 Finishing the installation ...
Page 69: ...Chapter 5 Troubleshooting ...
Page 71: ...Troubleshooting 59 ...