SE-601 Series Electrocardiograph User Manual EMC Information
- 73 -
Electromagnetic Immunity
Guidance and manufacture's declaration
– electromagnetic immunity
SE-601 series electrocardiograph
is intended for use in the electromagnetic environment
specified below. The customer or the user of SE-601 series electrocardiograph
should assure
that it is used in such an environment.
Immunity test
IEC/EN 60601 test
level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC/EN 61000-4-4
2 kV for power
supply lines
2 kV for power supply
lines
Mains
power
quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC/EN 61000-4-5
1 kV line to line
2 kV line to ground
1 kV line to line
2 kV line to ground
Mains
power
quality
should be that of a
typical commercial or
hospital environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC/EN 61000-4-8
30 A/m
30 A/m
Power
frequency
magnetic fields should
be
at
levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions
and
voltage
variations
on power supply
input lines
IEC/EN 61000-4-11
0 % U
T;
0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T;
0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Mains
power
quality
should be that of a
typical commercial or
hospital environment. If
the user of SE-601
series