SE-601 Series Electrocardiograph User Manual EMC Information
- 72 -
Appendix 2 EMC Information
Electromagnetic Emissions
Guidance and manufacture's declaration
– electromagnetic emission
SE-601 series electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or the user of SE-601 series electrocardiograph
should assure
that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
SE-601 series electrocardiograph
uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class A
SE-601 series electrocardiograph
is suitable
for use in all establishments, other than
domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Complies