SE-601 Series Electrocardiograph User Manual Technical Specifications
- 67 -
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC 60601-1:2005/A1:2012
EN 60601-1:2006/A1:2013
IEC 60601-1-2:2014
EN 60601-1-2:2015
IEC/EN 60601-2-25
Anti-electric-shock type:
Class І with internal power supply
Anti-electric-shock degree:
CF type with defibrillation-proof
Degree of protection against
harmful ingress of water:
Ordinary equipment (Sealed equipment without liquid proof)
Disinfection/sterilization
method:
Refer to Chapter 13 for details
Degree
of
safety
of
application in the presence
of flammable gas:
Equipment not suitable for use in the presence of flammable
gas
Working mode:
Continuous operation
EMC:
CISPR 11 Group 1, Class A
Patient
Leakage
Current:
NC
<10
μA (AC) / <10μA (DC)
SFC
<50μA (AC) / <50μA (DC)
Patient
Auxiliary
Current:
NC
<10μA (AC) / <10μA (DC)
SFC
<50μA (AC) / <50μA (DC)