SE-601 Series Electrocardiograph User Manual Operation Preparations
- 24 -
Snap/Banana Socket Adapters
Disposable Electrode (Snap Style)
Disposable Electrode Connection (Snap Style)
1)
Align all lead wires of the patient cable to avoid twisting, and cnnect Snap/Banana
Socket Adapters to connector of patient cable.
2)
Clean the electrode areas on the body surface with 75% alcohol.
3)
Attach the disposable electrodes to the electrode positions on the body surface.
4)
Connect Snap/Banana Socket Adapters to the disposable electrodes.
3.4 Inspection Before Power-On
In order to avoid safety hazards and get good ECG records, the following inspection procedures
are recommended before operation.
WARNING
The electrocardiograph is intended to be used by qualified physicians or personnel
professionally trained, and they should be familiar with the contents of this user manual
before operation.
1)
Environment
:
♦
Make sure that there is no electromagnetic interference source around the equipment,
especially large medical electrical equipment such as electrosurgical equipment,
radiological equipment, magnetic resonance imaging equipment etc. Turn off these
devices when necessary.
♦
Keep the examination room warm to avoid muscle tremor voltages in ECG signals
caused by cold.
2)
Power Supply:
♦
If the mains supply is used, please check whether the power cord is connected to the
unit well. The grounded three-slot outlet should be used.