SE-601 Series Electrocardiograph User Manual Technical Specifications
- 70 -
Minimum Amplitude:
20 μVp-p
Noise:
≤12.5µVp-p
Multichannel crosstalk
≤0.5mm
Filter
AC Filter:
On /Off
DFT Filter:
SE-601A:
0.05Hz/0.15Hz/0.25Hz/0.32Hz/0.5Hz/0.67Hz
SE-601B/C:
0.01Hz/0.05Hz/0.15Hz/0.25Hz/0.32Hz/0.5Hz/0.67Hz
EMG Filter:
25Hz/35Hz/45Hz/OFF
LOWPASS Filter:
SE-601A:
150Hz/100Hz/75Hz
SE-601B/C:
300Hz/270Hz/150Hz/100Hz/75Hz
CMRR
SE-601A:
≥115dB (AC OFF)
SE-601B/C:
≥123dB (AC OFF)
Sampling Frequency
SE-601A:
1,000/sec/channel
SE-601B/C:
16, 000/sec/channel
Pacemaker Detection
Amplitude
SE-601A:
±2 mV to ±700 mV
SE-601B/C: ±750uV to ±700 mV
Width
SE-601A:
0.1ms to 2.0 ms
SE-601B/C: 50
μs to 2.0 ms
Sampling Frequency
SE-601A:
10,000/sec/channel
SE-601B/C: 16,000/sec/channel
External Input/Output (Configurable)
Input
≥100kΩ; Sensitivity 10mm/V±5%;
Single ended
Output
≤100Ω; Sensitivity 1V/mV±5%;
Single ended
WIFI (Configurable)