4. After loosening the four bolt of "B", separate the existing battery pack lifted by
a hand.
5. Please remove the "C" connector from the connector housing "J1", ahead of
installing a new battery pack.
6. After removing the "C" connector from the connector housing "J1", Install a new
battery pack. Afterwards, fix the battery pack tightly with the bolts in the reverse
C
(J1)
Warning
Whenever installing a new battery pack after separating the existing device, the
"C" connector must be removed from connector housing of "J1" in advance. In
case of installation the battery pack without removing the "C" connector from
the connector housing "J1", it is able to damage the equipment by causing a
spark. If this case occurs with a equipment malfunction, please ask to the
retailer or manufacturer.