Calibration Mode
Because this equipment uses battery power source, long-time use of battery can
make some deviation of kV or mA value. Therefore once in year, should be
measure the kV or mA periodically by kV or mA Meter. If deviation is found,
calibrate it by using one of below Mode.
1. To access the Calibration Mode, turn off the equipment and turn on again while
pressing the 1 & 4 key together on the OP panel.
2. Select the Mode by kV key(UP/DOWN).
3. After selecting the Mode, adjust the deviation values by mAs key(UP/DOWN).
4. If calibration is finished, press 'APR Store key' for 2 second for storing new
calibrated values.
5. Turn off the equipment and turn on again.
6. Use the equipment.
Section
Setting
Function
Description
kV
kV mAs
Mode for adjust the kV
deviation.
Adjust from -5 to 5 by
using the mAs key
(UP/DOWN).
mA
kV mAs
Mode for adjust the mA
deviation.
Adjust from -5 to 5 by
using the mAs key
(UP/DOWN).
Summary of Contents for ANYPLX 3-BT
Page 4: ...Diagnostic X ray Unit USER MANUAL 4 Section 1 General Information ...
Page 22: ...Diagnostic X ray Unit USER MANUAL 22 n Rating Chart ...
Page 23: ...Diagnostic X ray Unit USER MANUAL 23 ...
Page 24: ...Diagnostic X ray Unit USER MANUAL 24 Section 2 Operation ...
Page 33: ...Diagnostic X ray Unit USER MANUAL 33 Section 3 Service Manual ...
Page 45: ...Diagnostic X ray Unit USER MANUAL 45 Section 4 Schematics ...
Page 46: ...Diagnostic X ray Unit USER MANUAL 46 Block Diagram ...
Page 47: ...Diagnostic X ray Unit USER MANUAL 47 Power Driver Board ...
Page 48: ...Diagnostic X ray Unit USER MANUAL 48 Main Board CPU ...
Page 49: ...Diagnostic X ray Unit USER MANUAL 49 Main Board EP_IP FB ...